| Title |
|
Ventricular fibrillation episodes related to managed ventricular pacing algorithm in a patient with mitral valve prolapse - "Arrhythmic MVP due to MVP". |
|
|
| Publication Date |
|
| Author(s) |
|
| Subject |
|
Managed ventricular pacing |
|
|
| Type of document |
|
| OrcId |
|
#PLACEHOLDER_PARENT_METADATA_VALUE# |
|
|
#PLACEHOLDER_PARENT_METADATA_VALUE# |
|
|
#PLACEHOLDER_PARENT_METADATA_VALUE# |
|
|
#PLACEHOLDER_PARENT_METADATA_VALUE# |
|
|
#PLACEHOLDER_PARENT_METADATA_VALUE# |
|
|
| DOI |
|
10.1016/j.jelectrocard.2023.03.008 |
|
|
| Abstract |
|
We report a case of an implant cardiac defibrillator (ICD) patient who an ICD implanted for ventricular fibrillation (VF) related to mitral valve prolapse. He has 2 episodes of VF in his device lives. First episode of VF in year 2016 was initiated after a pause related to the MVP™ algorithm with a critically timed PVC. MVP™ was turned off which prevented further VF episodes. However, MVP™ was turned back on at the device replacement in 2018. A second VF episode developed with similar mechanism in 2021 and MVP™ was subsequently turned off with no further VF recorded. This case highlights the importance of recognizing the mechanism of initiation of tachy-arrhythmia episodes and serves as an important reminder regarding optimization of device settings at the time of replacement. |
|
|
| Link |
|
| Citation |
|
Journal of Electrocardiology 2023; 79 |
|
|
| Jornal Title |
|
Journal of Electrocardiology |
|
|
| ISSN |
|