Please use this identifier to cite or link to this item: https://ahro.austin.org.au/austinjspui/handle/1/25271
Title: Precision of the Hologic Horizon A dual energy X-ray absorptiometry in the assessment of body composition.
Austin Authors: Cheung, Yee-Ming Melody ;Roff, G;Grossmann, Mathis 
Affiliation: Medicine (University of Melbourne)
Endocrinology
Issue Date: Nov-2020
Date: 2020-10-29
Publication information: Obesity Research & Clinical Practice 2020; 14(6): 514-518
Abstract: As the number of individuals with obesity increases, so too will the demand for a precise, cost-effective and safe method to better risk-stratify this heterogenous population. Visceral adipose tissue (VAT) area, a body composition measurement has been associated with the many facets of the metabolic syndrome. To determine the precision of dual energy X-ray absorptiometry (DXA) in providing body composition measurements, including VAT area, using the Hologic Horizon A scanner. Thirty participants (mean age 64 years, 47% male), underwent two sequential body composition scans. The coefficient of variation (CV) and least significant change (LSC) were calculated for android/gynoid ratio, appendicular lean/height2, lean/height2, total body fat and lean mass, total body mass, and subcutaneous adipose tissue (SAT) and VAT area. Participants were stratified according to age (<70 or ≥70 years), body mass index (BMI) (≤30 kg/m2 or >30 kg/m2) and sex (male or female). The CV was calculated for each parameter and then analysed for differences between groups. The CV for all parameters was <3% with the exception of SAT (3.86%). The CV of VAT area was 2.63% with a LSC at 95% confidence of 7.28%. Apart from the CV differences between males and females for android/gynoid ratio (male: 3.56% vs. female: 1.74%, p = 0.01) and SAT area (5.04% vs. 2.46%, p = 0.01), there were no significant differences identified between the calculated CV for all other body composition parameters when participants were stratified by age, BMI and sex. DXA scanning, using the Hologic Horizon A system, is capable of providing precise body composition measurements.
URI: https://ahro.austin.org.au/austinjspui/handle/1/25271
DOI: 10.1016/j.orcp.2020.10.005
Journal: Obesity Research & Clinical Practice
PubMed URL: 33132075
ISSN: 1871-403X
Type: Journal Article
Subjects: Body composition
Dual-energy X-ray absorptiometry
Metabolic syndrome
Precision
Visceral fat
Appears in Collections:Journal articles

Show full item record

Page view(s)

48
checked on Jul 22, 2024

Google ScholarTM

Check


Items in AHRO are protected by copyright, with all rights reserved, unless otherwise indicated.