Please use this identifier to cite or link to this item:
https://ahro.austin.org.au/austinjspui/handle/1/17984
Title: | High dose rate brachytherapy source measurement intercomparison. | Austin Authors: | Poder, Joel;Smith, Ryan L;Shelton, Nikki ;Whitaker, May;Butler, Duncan;Haworth, Annette | Affiliation: | Radiation Oncology, Chris O'Brien Lifehouse, Camperdown, NSW, Australia Centre of Medical Radiation Physics, University of Wollongong, Wollongong, NSW, Australia Alfred Health Radiation Oncology, The Alfred Hospital, Melbourne, Victoria, Australia School of Applied Sciences, RMIT University, Melbourne, Victoria, Australia Olivia Newton-John Cancer Wellness and Research Centre, Austin Health, Heidelberg, Victoria, Australia Australian Radiation Protection and Nuclear Safety Agency, Yallambie, Victoria, Australia Physical Sciences, Peter MacCallum Cancer Centre, Melbourne, Victoria, Australia School of Physics, University of Sydney, Camperdown, NSW, Australia |
Issue Date: | Jun-2017 | Date: | 2017-03-24 | Publication information: | Australasian physical & engineering sciences in medicine 2017; 40(2): 377-383 | Abstract: | This work presents a comparison of air kerma rate (AKR) measurements performed by multiple radiotherapy centres for a single HDR 192Ir source. Two separate groups (consisting of 15 centres) performed AKR measurements at one of two host centres in Australia. Each group travelled to one of the host centres and measured the AKR of a single 192Ir source using their own equipment and local protocols. Results were compared to the 192Ir source calibration certificate provided by the manufacturer by means of a ratio of measured to certified AKR. The comparisons showed remarkably consistent results with the maximum deviation in measurement from the decay-corrected source certificate value being 1.1%. The maximum percentage difference between any two measurements was less than 2%. The comparisons demonstrated the consistency of well-chambers used for 192Ir AKR measurements in Australia, despite the lack of a local calibration service, and served as a valuable focal point for the exchange of ideas and dosimetry methods. | URI: | https://ahro.austin.org.au/austinjspui/handle/1/17984 | DOI: | 10.1007/s13246-017-0542-6 | ORCID: | 0000-0003-3018-6750 | Journal: | Australasian physical & engineering sciences in medicine | PubMed URL: | 28342026 | Type: | Journal Article | Subjects: | Brachytherapy Calibration Intercomparison Well-chamber |
Appears in Collections: | Journal articles |
Show full item record
Items in AHRO are protected by copyright, with all rights reserved, unless otherwise indicated.