An investigation of the basic physics of irrigation in urology and the role of automated pump irrigation in cystoscopy.
Publication Date
2012-05-15
Author(s)
Chang, Dwayne
Manecksha, Rustom P
Syrrakos, Konstantinos
Lawrentschuk, Nathan
Type of document
Journal Article
DOI
10.1100/2012/476759
Abstract
To investigate the effects of height, external pressure, and bladder fullness on the flow rate in continuous, non-continuous cystoscopy and the automated irrigation fluid pumping system (AIFPS).Each experiment had two 2-litre 0.9% saline bags connected to a continuous, non-continuous cystoscope or AIFPS via irrigation tubing. Other equipment included height-adjustable drip poles, uroflowmetry devices, and model bladders.In Experiment 1, saline bags were elevated to measure the increment in flow rate. In Experiment 2, saline bags were placed under external pressures to evaluate the effect on flow rate. In Experiment 3, flow rate changes in response to variable bladder fullness were measured.Elevating saline bags caused an increase in flow rates, however the increment slowed down beyond a height of 80 cm. Increase in external pressure on saline bags elevated flow rates, but inconsistently. A fuller bladder led to a decrease in flow rates. In all experiments, the AIFPS posted consistent flow rates.Traditional irrigation systems were susceptible to changes in height of irrigation solution, external pressure application, and bladder fullness thus creating inconsistent flow rates. The AIFPS produced consistent flow rates and was not affected by any of the factors investigated in the study.
Link
Citation
Thescientificworldjournal 2012; 2012(): 476759
Jornal Title
TheScientificWorldJournal

Files:

NameSizeformatDescriptionLink
22666130.pdf 1472.939 KB application/pdf View document