Please use this identifier to cite or link to this item: https://ahro.austin.org.au/austinjspui/handle/1/10683
Title: The effect of three different miniaturized blood purification devices on plasma cytokine concentration in an ex vivo model of endotoxinemia.
Austin Authors: Haase, M;Bellomo, Rinaldo ;Baldwin, Ian C ;Haase-Fielitz, Anja;Storr, M;Krause, B;Boyce, Neil;Svobodova, S;Li, W;Bagshaw, Sean M;Warrillow, Stephen J ;Langenberg, Christoph;Morgera, Stanislao
Affiliation: Intensive Care Research, Austin Health, University of Melbourne, Australia
Issue Date: 1-Aug-2008
Publication information: The International Journal of Artificial Organs; 31(8): 722-9
Abstract: A novel type of adsorptive plasma filtering device (ETX-A) capable of removing endotoxin from blood in a single step has recently been developed using nanotechnology.In a miniaturized, ex vivo model of extracorporeal circuits, we tested the capacity to reduce plasma cytokine concentration of ETX-A filters in comparison to standard high-flux (HF) filters, high cut-off (HCO) filters and a control. Blood from six healthy volunteers was spiked with endotoxin and then circulated through closed (ETX-A, control) or open (HF, HCO) circuits. Blood flow was set at 16 ml/min and filtration flow at 1 ml/min. Samples for measurement of IL-1ra and IL-6 were taken at baseline and at 4 hours.Compared to control (703.3 [850.6] pg/mL), in HCO (383.5 [1144.1] pg/mL) and ETX-A (490.1 [683.2] pg/mL) filters, plasma IL-1ra pooled pre- and postfilter concentrations were lower at the end of the experiment (P=0.002; P=0.050, respectively) whereas, in standard HF filters, IL-1ra concentration was higher than control. HCO showed a trend toward a reduced relative increase in IL-6 concentration from commencement to end of experiment compared to control (P=0.07). After pooling end-of-experiment plasma cytokine values of novel blood purification devices, we found HCO + ETX-A superior to H with regard to reduction of IL-1ra (-27.0 [-20.5]% vs. 8.1 [18.9]%; p<0.001) and IL-6 (-18.0 [38.3]% vs. -1.1 [24.3]%; P=0.050) compared to control.HCO and ETX-A appeared to significantly reduce plasma IL-1ra and, when combined, plasma IL-6 concentration as well. It appears desirable to manufacture full-size blood purification devices using this technology and to explore their effect on cytokine removal.
Gov't Doc #: 18825645
URI: https://ahro.austin.org.au/austinjspui/handle/1/10683
Journal: International Journal of Artificial Organs
URL: https://pubmed.ncbi.nlm.nih.gov/18825645
Type: Journal Article
Subjects: Adult
Endotoxemia.immunology.therapy
Equipment Design
Hemofiltration.instrumentation
Humans
Interleukin 1 Receptor Antagonist Protein.blood
Interleukin-6.blood
Lipopolysaccharides.blood
Male
Middle Aged
Miniaturization
Time Factors
Appears in Collections:Journal articles

Show full item record

Page view(s)

22
checked on Oct 4, 2024

Google ScholarTM

Check


Items in AHRO are protected by copyright, with all rights reserved, unless otherwise indicated.