Title: Non-invasive automated measurement of cardiac output during stable cardiac surgery using a fully integrated differential CO(2) Fick method.
Austin Authors: Peyton, Philip J ;Thompson, Daniel;Junor, Paul
Affiliation: Department of Anaesthesia, Austin Hospital, University of Melbourne, Melborne, Australia
Issue Date: 12-Jul-2008
Publication information: Journal of Clinical Monitoring and Computing 2008; 22(4): 285-92
Abstract: To re-evaluate the accuracy and precision of a non-invasive method for measurement of cardiac output based on the differential CO(2) Fick approach using an automated change in respiratory rate delivered by a ventilator under control by a prototype measurement system.Twenty-four patients during coronary artery bypass surgery, pre- and postcardiopulmonary bypass were recruited. After routine cannulation including pulmonary artery catheter, relaxant general anesthesia was induced. After hemodynamic and ventilatory stability were achieved, simultaneous paired measurements were made by the differential Fick method and by bolus thermodilution. Measurements were generated by inducing a change in respiratory rate by the ventilator under computer control. In Group 1, this involved an increase in respiratory rate from 8 to 12 breaths/min. In Group 2, this involved a decrease from 12 to 6 breaths/min.Nineteen measurements were made in each Group, 12 pre-CPB and 7 post-CPB. In Group 1 mean bias was -0.06 l/min, with a precision of agreement of 0.87 l/min, r = 0.91. In Group 2 (excluding one outlier) mean bias was -0.07 l/min, with a precision of 1.12 l/min, r = 0.71.Acceptable agreement with thermo- dilution during surgery was found, particularly where the ventilatory change involved an increase in respiratory rate from a lower baseline. This approach has potential to be readily integrated into modern anesthesia delivery platforms, allowing routine non-invasive cardiac output measurement.
Gov't Doc #: 18622583
URI: https://ahro.austin.org.au/austinjspui/handle/1/10638
DOI: 10.1007/s10877-008-9131-2
Journal: Journal of clinical monitoring and computing
URL: https://pubmed.ncbi.nlm.nih.gov/18622583
Type: Journal Article
Subjects: Adult
Aged
Breath Tests.instrumentation
Carbon Dioxide.analysis
Cardiac Output
Cardiac Surgical Procedures.instrumentation.methods
Female
Humans
Intraoperative Care.methods
Male
Middle Aged
Monitoring, Physiologic.instrumentation.methods
Reproducibility of Results
Sensitivity and Specificity
Surgery, Computer-Assisted.instrumentation.methods
Systems Integration
Appears in Collections:Journal articles

Files in This Item:
File SizeFormat 
18622583.pdf50.05 kBAdobe PDFView/Open
Show full item record

Page view(s)

30
checked on Nov 20, 2024

Download(s)

102
checked on Nov 20, 2024

Google ScholarTM

Check


Items in AHRO are protected by copyright, with all rights reserved, unless otherwise indicated.