Please use this identifier to cite or link to this item: https://ahro.austin.org.au/austinjspui/handle/1/10255
Title: A pilot assessment of the FloTrac cardiac output monitoring system.
Austin Authors: Opdam, Helen Ingrid;Wan, Li;Bellomo, Rinaldo 
Affiliation: Department of Intensive Care, Austin Hospital, VIC 3084, Heidelberg, Australia
Issue Date: 25-Oct-2006
Publication information: Intensive Care Medicine 2006; 33(2): 344-9
Abstract: To compare measurement of cardiac output (CO) by means of the FloTrac CO monitor with the pulmonary artery catheter (PAC).Prospective observational study.Intensive care unit of a tertiary hospital.Six post-operative cardiac surgery patients with existing arterial cannulas and PACs.Attachment of the FloTrac CO monitor and transducer to an existing arterial cannula. Simultaneous measurements of CO, indexed to body surface area (cardiac index, CI) by the FloTrac CO monitor and by either a bolus thermodilution or continuous CO PAC. Statistical analysis of observations.We performed CO measurements in six patients every 1-4 h after cardiac surgery. Comparison of all measurements showed a limited correlation for CI with the two devices (r (2)=0.1218, bias=0.21, 95% limits of agreement -0.81, 1.23). CI measurements obtained with the intermittent bolus PAC had better correlation with the FloTrac CI values (r (2) = 0.2693, bias=-0.0057, 95% limits of agreement -1.2042, 1.1929) than did those obtained with the continuous CO PAC (r (2)=0.0557, bias=0.2436, 95% limits of agreement -0.7350, 1.2222). When analysed according to heart rhythm, CI values measured during atrial pacing showed the best correlation (r (2)=0.377, bias=-0.0244, 95% limits of agreement -0.5226, 0.5714).CO measurements obtained using the FloTrac CO monitor show a limited correlation with those acquired using the PAC, relatively wide limits of agreement but no clear bias. Further evaluation is required before this device can be recommended for use in the clinical setting.
Gov't Doc #: 17063359
URI: https://ahro.austin.org.au/austinjspui/handle/1/10255
DOI: 10.1007/s00134-006-0410-4
Journal: Intensive Care Medicine
URL: https://pubmed.ncbi.nlm.nih.gov/17063359
Type: Journal Article
Subjects: Aged, 80 and over
Aortic Valve
Cardiac Output
Catheterization, Swan-Ganz
Coronary Artery Bypass
Evaluation Studies as Topic
Female
Heart Valves.surgery
Humans
Intensive Care Units
Male
Middle Aged
Mitral Valve
Monitoring, Physiologic.instrumentation
Postoperative Period
Appears in Collections:Journal articles

Files in This Item:
File SizeFormat 
17063359.pdf47.65 kBAdobe PDFView/Open
Show full item record

Page view(s)

36
checked on Nov 22, 2024

Download(s)

94
checked on Nov 22, 2024

Google ScholarTM

Check


Items in AHRO are protected by copyright, with all rights reserved, unless otherwise indicated.