Please use this identifier to cite or link to this item: https://ahro.austin.org.au/austinjspui/handle/1/30121
Title: Exclusion of EEG-based arousals in wake epochs of polysomnography leads to underestimation of the arousal index.
Austin Authors: Wilson, Danielle L ;Tolson, Julie ;Churchward, Thomas J ;Melehan, Kerri;O'Donoghue, Fergal J ;Ruehland, Warren R 
Affiliation: Respiratory and Sleep Medicine
Institute for Breathing and Sleep
Royal Prince Alfred Hospital, Sydney, Australia
Faculty of Medicine, University of Melbourne, Melbourne, Australia
Faculty of Medicine and Health, University of Sydney, Sydney, Australia
Issue Date: 1-May-2022
Publication information: Journal of Clinical Sleep Medicine 2022; 18(5): 1385-1393
Abstract: There is an internal contradiction in current American Academy of Sleep Medicine standards for arousal index (AI) calculation in polysomnography: Arousals in sleep and wake epochs are counted, but only sleep time is used in the denominator. This study aimed to investigate the impact of including arousals scored in wake epochs on the AI. We compared AIs including (AIinc) vs excluding (AIexc) awake-epoch arousals from 100 consecutive polysomnograms conducted for investigation of possible obstructive sleep apnea. To determine the AI that most closely approximated "truth," AIinc and AIexc were compared to an AI calculated from continuous sleep analysis (AIcont) in a 20-polysomnogram subgroup of patients. The median (interquartile range) increase in AIinc was 5.2 events/h (3.5-8.1) vs AIexc (AIinc = 28.0 events/h [18.4-38.9] vs AIexc = 22.9 events/ h [13.1-31.3]), equating to an increase of 25.3% (15.6-40.8). As the AI increased, the difference increased (P < .001), with decreasing sleep efficiency and an increasing apnea-hypopnea index as the strongest predictors of the difference between AIexc and AIinc. The absolute AIexc-AIcont difference (7.7 events/h [5.1-13.6]) was significantly greater than the AIinc-AIcont difference (1.2 events/h [0.6-5.7]; z = -3.099; P = .002). There was a notable increase in AI when we included wake-epoch arousals, particularly in patients with more severe obstructive sleep apnea or reduced sleep efficiency. However, the AI including wake-epoch arousals best matched the "true" continuous sleep-scoring AI. Our study informs clinical and research practice, highlights epoch scoring pitfalls, and supports the current American Academy of Sleep Medicine standard arousal reporting approach for future standards. Wilson DL, Tolson J, Churchward TJ, Melehan K, O'Donoghue FJ, Ruehland WR. Exclusion of EEG-based arousals in wake epochs of polysomnography leads to underestimation of the arousal index. J Clin Sleep Med. 2022;18(5):1385-1393.
URI: https://ahro.austin.org.au/austinjspui/handle/1/30121
DOI: 10.5664/jcsm.9878
ORCID: 0000-0001-6754-7359
0000-0002-7002-2657
0000-0003-1471-9318
0000-0001-5099-3184
0000-0001-9626-7460
Journal: Journal of Clinical Sleep Medicine : JCSM : Official Publication of the American Academy of Sleep Medicine
PubMed URL: 35022129
PubMed URL: https://pubmed.ncbi.nlm.nih.gov/35022129/
Type: Journal Article
Subjects: analysis
arousal
continuous
epoch
polysomnography
wake
Appears in Collections:Journal articles

Show full item record

Page view(s)

74
checked on Dec 24, 2024

Google ScholarTM

Check


Items in AHRO are protected by copyright, with all rights reserved, unless otherwise indicated.