Please use this identifier to cite or link to this item: https://ahro.austin.org.au/austinjspui/handle/1/22832
Full metadata record
DC FieldValueLanguage
dc.contributor.authorHardcastle, Nicholas-
dc.contributor.authorKron, Tomas-
dc.contributor.authorCook, Olivia-
dc.contributor.authorLehmann, Joerg-
dc.contributor.authorMitchell, Paul L R-
dc.contributor.authorSiva, Shankar-
dc.date2020-04-
dc.date.accessioned2020-03-23T22:10:37Z-
dc.date.available2020-03-23T22:10:37Z-
dc.date.issued2020-03-16-
dc.identifier.citationPhysica Medica : PM 2020; 72: 16-21-
dc.identifier.urihttps://ahro.austin.org.au/austinjspui/handle/1/22832-
dc.description.abstractStereotactic ablative body radiotherapy (SABR) in multi-centre trials requires rigorous quality assurance to ensure safe and consistent treatment for all trial participants. We report results of vertebral SABR dosimetry credentialing for the ALTG/TROG NIVORAD trial. Centres with a previous SABR site visit performed axial film measurement of the benchmarking vertebral plan in a local phantom and submitted radiochromic film images for analysis. Remaining centres had on-site review of SABR processes and axial film measurement of the vertebral benchmarking plan. Films were analysed for dosimetric and positional accuracy: gamma analysis (>90% passing 2%/2mm/10% threshold) and ≤ 1 mm positional accuracy at target-cord interface was required. 19 centres were credentialed; 11 had on-site measurement. Delivery devices included linear accelerator, TomoTherapy and CyberKnife systems. Five centres did not achieve 90% gamma passing rate. Of these, three were out of tolerance (OOT) in low (<5Gy) dose regions and > 80% passing rate and deemed acceptable. Two were OOT over the full dose range: one elected not to remeasure; the other also had positional discrepancy greater than 1 mm and repeat measurement with a new plan was in tolerance. The original OOT was attributed to inappropriate MLC constraints. All centres delivered planned target-cord dose gradient within 1 mm. Credentialing measurements for vertebral SABR in a multi-centre trial showed although the majority of centres delivered accurate vertebral SABR, there is high value in independent audit measurements. One centre with inappropriate MLC settings was detected, which may have resulted in delivery of clinically unacceptable vertebral SABR plans.-
dc.language.isoeng-
dc.subjectClinical trial-
dc.subjectDosimetry-
dc.subjectImmunotherapy-
dc.subjectLung cancer-
dc.subjectSABR-
dc.subjectSBRT-
dc.subjectVertebral-
dc.titleCredentialing of vertebral stereotactic ablative body radiotherapy in a multi-centre trial.-
dc.typeJournal Article-
dc.identifier.journaltitlePhysica Medica : PM-
dc.identifier.affiliationRadiation Oncology, Peter MacCallum Cancer Centre, VIC, Australiaen
dc.identifier.affiliationDepartment of Radiation Oncology, Calvary Mater Newcastle, NSW, Australiaen
dc.identifier.affiliationOlivia Newton-John Cancer Wellness and Research Centre, Austin Health, Heidelberg, Victoria, Australiaen
dc.identifier.affiliationDepartment of Medical Oncology, Austin Health, Heidelberg, Victoria, Australiaen
dc.identifier.affiliationInstitute of Medical Physics, The University of Sydney, NSW, Australiaen
dc.identifier.affiliationTrans Tasman Radiation Oncology Group, NSW, Australiaen
dc.identifier.affiliationPhysical Sciences, Peter MacCallum Cancer Centre, VIC, Australiaen
dc.identifier.affiliationCentre for Medical Radiation Physics, University of Wollongong, NSW, Australiaen
dc.identifier.affiliationSir Peter MacCallum Cancer Centre, University of Melbourne, VIC, Australiaen
dc.identifier.affiliationSchool of Mathematical and Physical Sciences, The University of Newcastle, NSW, Australiaen
dc.identifier.doi10.1016/j.ejmp.2020.03.004-
dc.identifier.pubmedid32193090-
dc.type.austinJournal Article-
local.name.researcherMitchell, Paul L R
item.openairetypeJournal Article-
item.cerifentitytypePublications-
item.grantfulltextnone-
item.fulltextNo Fulltext-
item.openairecristypehttp://purl.org/coar/resource_type/c_18cf-
item.languageiso639-1en-
crisitem.author.deptMedical Oncology-
crisitem.author.deptOlivia Newton-John Cancer Wellness and Research Centre-
Appears in Collections:Journal articles
Show simple item record

Page view(s)

42
checked on Nov 25, 2024

Google ScholarTM

Check


Items in AHRO are protected by copyright, with all rights reserved, unless otherwise indicated.