Please use this identifier to cite or link to this item: https://ahro.austin.org.au/austinjspui/handle/1/10709
Full metadata record
DC FieldValueLanguage
dc.contributor.authorNaka, Toshioen
dc.contributor.authorEgi, Moritokien
dc.contributor.authorBellomo, Rinaldoen
dc.contributor.authorBaldwin, Ian Cen
dc.contributor.authorFealy, Nigel Gen
dc.contributor.authorWan, Lien
dc.date.accessioned2015-05-16T00:14:49Z
dc.date.available2015-05-16T00:14:49Z
dc.date.issued2008-10-01en
dc.identifier.citationThe International Journal of Artificial Organs; 31(10): 905-9en
dc.identifier.govdoc19009509en
dc.identifier.otherPUBMEDen
dc.identifier.urihttps://ahro.austin.org.au/austinjspui/handle/1/10709en
dc.description.abstractTo assess the resistance posed by double-lumen vascular access dialysis catheters at low and high blood flow.Controlled ex vivo study Setting: ICU Laboratory of tertiary hospital.Eleven proprietary vascular access catheters for continuous renal replacement therapy.Heparinized spent red cells diluted in polygeline solution were pumped using the Aquarius hemofiltration machine (Edwards Life Sciences, Sydney, NSW, Australia) and its standard circuit through several vascular access catheters. Blood flow was increased and then decreased in steps of 50 ml/min (50, 150, 200, 250 and 300 ml/min) while catheter outflow and inflow pressures were recorded. The pressure-flow relationship (hydraulic resistance) of each catheter was then calculated. Study catheters were divided into two groups according to their internal diameter (large gauge vs. smaller gauge) or length (long or short). Hydraulic resistances were compared between the groups.Different double lumen catheters posed clearly different resistances to flow. For all groups of catheters, there was a linear relationship between pressure and flow. No statistically significant difference between short and long catheters could be demonstrated (p=0.715). On the other hand, larger gauge catheters (13 Fr or greater) had significantly lower resistances than smaller gauge (<13 Fr) catheters (p=0.0062). Furthermore, all larger gauge catheters had resistances lower than 0.430 mmHg/ml/min, while all smaller gauge catheters had resistances greater than 0.490 mmHg/ml/min.Commercial double-lumen dialysis catheters have variable resistance to blood flow under standard ex vivo conditions. Although both length and internal diameter varied, internal diameter had a dominant effect on resistance. This information might be useful to clinicians in guiding their choice of catheters for clinical use.en
dc.language.isoenen
dc.subject.otherBlood Flow Velocity.physiologyen
dc.subject.otherBlood Pressure.physiologyen
dc.subject.otherCatheters, Indwelling.adverse effectsen
dc.subject.otherEquipment Designen
dc.subject.otherErythrocytesen
dc.subject.otherHeparinen
dc.subject.otherHumansen
dc.subject.otherPressureen
dc.subject.otherRenal Replacement Therapy.methodsen
dc.subject.otherVascular Resistance.physiologyen
dc.titleResistance of vascular access catheters for continuous renal replacement therapy: An ex vivo evaluation.en
dc.typeJournal Articleen
dc.identifier.journaltitleInternational Journal of Artificial Organsen
dc.identifier.affiliationDepartment of Intensive Care and Department of Medicine, Austin Hospital, Melbourne - Australiaen
dc.description.pages905-9en
dc.relation.urlhttps://pubmed.ncbi.nlm.nih.gov/19009509en
dc.type.austinJournal Articleen
local.name.researcherBaldwin, Ian C
item.openairetypeJournal Article-
item.cerifentitytypePublications-
item.grantfulltextnone-
item.fulltextNo Fulltext-
item.openairecristypehttp://purl.org/coar/resource_type/c_18cf-
item.languageiso639-1en-
crisitem.author.deptIntensive Care-
crisitem.author.deptData Analytics Research and Evaluation (DARE) Centre-
crisitem.author.deptIntensive Care-
crisitem.author.deptIntensive Care-
Appears in Collections:Journal articles
Show simple item record

Page view(s)

70
checked on Nov 22, 2024

Google ScholarTM

Check


Items in AHRO are protected by copyright, with all rights reserved, unless otherwise indicated.