Please use this identifier to cite or link to this item: https://ahro.austin.org.au/austinjspui/handle/1/13148
Full metadata record
DC FieldValueLanguage
dc.contributor.authorDudson, Jen
dc.date.accessioned2015-05-16T02:56:10Z
dc.date.available2015-05-16T02:56:10Z
dc.date.issued1994-12-01en
dc.identifier.citationAustralasian Physical & Engineering Sciences in Medicine / Supported By the Australasian College of Physical Scientists in Medicine and the Australasian Association of Physical Sciences in Medicine; 17(4): 211-6en
dc.identifier.govdoc7872903en
dc.identifier.otherPUBMEDen
dc.identifier.urihttp://ahro.austin.org.au/austinjspui/handle/1/13148en
dc.description.abstractA recently introduced Agfa film-screen combination used for mammography was compared with a widely used Kodak film-screen combination. This comparison involved evaluating the image quality and calculating a reference mean glandular dose for a breast phantom recommended by the ACPSEM. The image quality and dose were evaluated for a grid and non-grid technique as well as a range of kilovoltages. For the large area test objects in the phantom, the Agfa combination performed better. This reflected the high contrast capability of this film-screen combination. For the Kodak combination a better image score was achieved for the small area objects. These objects are indicative of the modulation transfer function and resolution characteristics of the film-screen combination. The reference mean glandular dose showed that at least 20% less dose was required for the Agfa combination. Although the Agfa combination offered improved contrast and dose saving, the Kodak combination was preferred because of the better representation of the small area objects which are an important element for early detection of breast cancer.en
dc.language.isoenen
dc.subject.otherEquipment Designen
dc.subject.otherFemaleen
dc.subject.otherHumansen
dc.subject.otherMammography.instrumentationen
dc.subject.otherModels, Structuralen
dc.subject.otherRadiographic Image Enhancement.methodsen
dc.subject.otherX-Ray Filmen
dc.titleA comparison of an Agfa and Kodak film-screen combination for mammography.en
dc.typeJournal Articleen
dc.identifier.journaltitleAustralasian physical & engineering sciences in medicine / supported by the Australasian College of Physical Scientists in Medicine and the Australasian Association of Physical Sciences in Medicineen
dc.identifier.affiliationDepartment of Medical Physics, Austin Hospital, Heidelberg, Victoria, Australiaen
dc.description.pages211-6en
dc.relation.urlhttps://pubmed.ncbi.nlm.nih.gov/7872903en
dc.type.austinJournal Articleen
item.openairecristypehttp://purl.org/coar/resource_type/c_18cf-
item.fulltextNo Fulltext-
item.cerifentitytypePublications-
item.grantfulltextnone-
item.languageiso639-1en-
item.openairetypeJournal Article-
Appears in Collections:Journal articles
Show simple item record

Google ScholarTM

Check


Items in AHRO are protected by copyright, with all rights reserved, unless otherwise indicated.