dc.identifier.urihttps://ahro.austin.org.au/austinjspui/handle/1/9886en
dc.description.abstractTo test the ability of a novel super high-flux (SHF) membrane with a larger pore size to clear myoglobin from serum.The intensive care unit of a university teaching hospital.A patient with serotonin syndrome complicated by severe rhabodomyolysis and oliguric acute renal failure.Initially continuous veno-venous hemofiltration was performed at 2 l/hour ultrafiltration (UF) with a standard polysulphone 1.4 m2 membrane (cutoff point, 20 kDa), followed by continuous veno-venous hemofiltration with a SHF membrane (cutoff point, 100 kDa) at 2 l/hour UF, then at 3 l/hour UF and then at 4 l/hour UF, in an attempt to clear myoglobin.The myoglobin concentration in the ultrafiltrate at 2 l/hour exchange was at least five times greater with the SHF membrane than with the conventional membrane (>100,000 microg/l versus 23,003 microg/l). The sieving coefficients with the SHF membrane at 3 l/hour UF and 4 l/hour UF were 72.2% and 68.8%, respectively. The amount of myoglobin removed with the conventional membrane was 1.1 g/day compared with 4.4-5.1 g/day for the SHF membrane. The SHF membrane achieved a clearance of up to 56.4 l/day, and achieved a reduction in serum myoglobin concentration from >100,000 microg/l to 16,542 microg/l in 48 hours.SHF hemofiltration achieved a much greater clearance of myoglobin than conventional hemofiltration, and it may provide a potential modality for the treatment of myoglobinuric acute renal failure.en
dc.language.isoenen
dc.subject.otherAcute Kidney Injury.complicationsen
dc.subject.otherFemaleen
dc.subject.otherGlasgow Coma Scaleen
dc.subject.otherHemofiltration.instrumentation.methodsen
dc.subject.otherHospitals, Teachingen
dc.subject.otherHumansen
dc.subject.otherIntensive Care Unitsen
dc.subject.otherMiddle Ageden
dc.subject.otherMyoglobin.blooden
dc.subject.otherMyoglobinuria.therapyen
dc.subject.otherOliguria.complicationsen
dc.subject.otherRhabdomyolysis.therapyen
dc.subject.otherSerotonin Syndrome.complications.diagnosis.therapyen
dc.subject.otherTime Factorsen
dc.titleMyoglobin clearance by super high-flux hemofiltration in a case of severe rhabdomyolysis: a case report.en
dc.typeJournal Articleen
dc.identifier.journaltitleCritical Careen
dc.identifier.affiliationDepartment of Intensive Care, Melbourne University, Austin Hospital, Melbourne, Australiaen
dc.identifier.doi10.1186/cc3034en
dc.description.pagesR90-5en
dc.relation.urlhttps://pubmed.ncbi.nlm.nih.gov/15774055en
dc.type.austinJournal Articleen
local.name.researcherBaldwin, Ian C
item.openairetypeJournal Article-
item.openairecristypehttp://purl.org/coar/resource_type/c_18cf-
item.grantfulltextopen-
item.fulltextWith Fulltext-
item.cerifentitytypePublications-
item.languageiso639-1en-
crisitem.author.deptIntensive Care-
crisitem.author.deptIntensive Care-
crisitem.author.deptIntensive Care-
crisitem.author.deptIntensive Care-
crisitem.author.deptData Analytics Research and Evaluation (DARE) Centre-
Appears in Collections:Journal articles
Files in This Item:
File Description SizeFormat 
15774055.pdf444.63 kBAdobe PDFThumbnail
View/Open
Show simple item record

Page view(s)

10
checked on Mar 27, 2024

Download(s)

80
checked on Mar 27, 2024

Google ScholarTM

Check


Items in AHRO are protected by copyright, with all rights reserved, unless otherwise indicated.