Title: Validation of a solid-phase extraction high-performance liquid chromatographic assay for doxazosin.
Authors: Jackman, G P;Colagrande, F;Louis, William J
Affiliation: Department of Clinical Pharmacology and Therapeutics, Austin Hospital, Heidelberg, Victoria, Australia
Issue Date: 3-May-1991
Citation: Journal of Chromatography; 566(1): 234-8
Abstract: The determination of doxazosin by high-performance liquid chromatography with fluorescence detection is described. Propanolol was used as the internal standard. Plasma samples were treated with methanol to precipitate the proteins. Doxazosin was isolated with C18 reversed-phase extraction columns. The determination limit is 1 ng/ml of plasma, while the extraction columns can be reused frequently. The method is applied to clinical trial samples.
Internal ID Number: 1679435
URI: http://ahro.austin.org.au/austinjspui/handle/1/10184
URL: http://www.ncbi.nlm.nih.gov/pubmed/1679435
Type: Journal Article
Subjects: Adrenergic alpha-Antagonists.blood
Chromatography, High Pressure Liquid.methods.standards.statistics & numerical data
Doxazosin
Humans
Hydrogen-Ion Concentration
Methanol
Prazosin.analogs & derivatives.blood
Propranolol
Appears in Collections:Journal articles